Catálogo de publicaciones - revistas
Biomedical Physics & Engineering Express
Resumen/Descripción – provisto por la editorial en inglés
A broad, inclusive, rapid review journal devoted to publishing new research in all areas of biomedical engineering, biophysics and medical physics, with a special emphasis on interdisciplinary work between these fields.Palabras clave – provistas por la editorial
No disponibles.
Disponibilidad
Institución detectada | Período | Navegá | Descargá | Solicitá |
---|---|---|---|---|
No detectada | desde jun. 2015 / hasta dic. 2023 | IOPScience |
Información
Tipo de recurso:
revistas
ISSN electrónico
2057-1976
Editor responsable
IOP Publishing (IOP)
País de edición
Estados Unidos
Fecha de publicación
2015-
Cobertura temática
Tabla de contenidos
Biological impact of dosimetric perturbations of a fiducial marker and the daily number of fields in proton therapy for prostate cancer
Tatstuhiko Sagara; Takahiro Kato; Masao Murakami
<jats:title>Abstract</jats:title> <jats:p>The purpose of this study was to estimate the biological impact of dosimetric perturbations of a fiducial marker and the daily number of fields in proton therapy for prostate cancer. Using a linear-quadratic model, normalized total doses (NTDs) of points where deposited dose was reduced from the prescribed dose by dosimetric perturbation of a fiducial marker were calculated in two hypothetical prostate cancer treatment schedules: a) irradiation of both parallel-opposed lateral fields and b) irradiation of alternate field in each daily treatment. The impact of hypofractionation and sublethal damage repair between irradiation on NTD was also estimated. The NTD of two fields/day schedule becomes lower than that of one field/day schedule. The difference becomes larger as dose reduction from one of two fields becomes more enhanced. The NTD reduction from the total dose in the two fields/day schedule is largest (30% of total dose) where the dose from one beam is completely lost by a fiducial marker. In contrast, the NTD reduction from the total dose in the one field/day schedule is largest (9% of total dose) where the half dose from one beam is decreased by a fiducial marker. In addition, the NTD reduction becomes larger as the fractional dose increases in a hypofractionated regimen, and when the effect of sublethal damage repair was incorporated. These influences become significant in prostate cancer since the radiobiological sensitivity <jats:italic>α</jats:italic>/<jats:italic>β</jats:italic> of prostate cancer is lower than other cancer types and normal tissues late complication. Treating with one alternate field in a daily treatment can improve a deteriorating treatment effect by dosimetric distortion of a fiducial marker in prostate cancer treatment. However, the choice of the number of beams in a fraction must also be determined by considering the sparing of normal tissues and patient-specific status.</jats:p>
Palabras clave: General Nursing.
Pp. 025007
EEG signal analysis during Ishihara’s test in subjects with normal vision and color vision deficiency
Ali Ekhlasi; Hessam Ahmadi; Amir Molavi; Mohammad Saadat Nia; Ali Motie Nasrabadi
<jats:title>Abstract</jats:title> <jats:p>Color Vision Deficiency (CVD) is one of the most common types of vision deficiency. People with CVD have difficulty seeing color spectra depending on what types of retina photoreceptors are impaired. In this paper, the Ishihara test with 38 plates was used to examine the Electroencephalogram (EEG) of ten subjects with CVD plus ten healthy individuals. The recording was performed according to the 10–20 international system. The C-based software was programmed so that subjects could select the number or path in each test plate in the software options while recording EEG. Frequency features in different frequency bands were extracted from the EEG signals of the two groups during the Ishihara test. Statistically significant differences (P < 0.05) between features were assessed by independent samples t-test with False Discovery Rate (FDR) correction. Also, the K-nearest neighbor classifier (KNN) was used to classify the two groups. The results revealed that the most significant difference between the two groups in the Ishihara test images occurred for the electrodes located in the right temporoparietal areas (P4 and T6) of the brain in the Delta, Theta, Beta1, and Beta2 frequency bands. The KNN classifier, using the signals that reported the greatest statistical difference between the two groups, showed that the two groups were distinguishable with 85.2% accuracy. In this way, images from the Ishihara test that would provide the most accurate classification were identified. In conclusion, this research provided new insights into EEG signals of subjects with CVD and healthy subjects based on the Ishihara color vision test.</jats:p>
Palabras clave: General Nursing.
Pp. 025008
Minimizing 4DCBCT imaging dose and scan time with Respiratory Motion Guided 4DCBCT: a pre-clinical investigation
Tess Reynolds; Praise Lim; Paul J Keall; Ricky O’Brien
<jats:title>Abstract</jats:title> <jats:p>Current conventional 4D Cone Beam Computed Tomography (4DCBCT) imaging is hampered by inconsistent patient breathing that leads to long scan times, reduced image quality and high imaging dose. To address these limitations, Respiratory Motion Guided 4D cone beam computed tomography (RMG-4DCBCT) uses mathematical optimization to adapt the gantry rotation speed and projection acquisition rate in real-time in response to changes in the patient’s breathing rate. Here, RMG-4DCBCT is implemented on an Elekta Synergy linear accelerator to determine the minimum achievable imaging dose. 8 patient-measured breathing traces were programmed into a 1D motion stage supporting a 3D-printed anthropomorphic thorax phantom. The respiratory phase and current gantry position were calculated in real-time with the RMG-4DCBCT software, which in turn modulated the gantry rotation speed and suppressed projection acquisition. Specifically, the effect of acquiring 20, 25, 30, 35 and 40 projections/respiratory phase bin RMG scans on scan time and image quality was assessed. Reconstructed image quality was assessed via the contrast-to-noise ratio (CNR) and the Edge Response Width (ERW) metrics. The performance of the system in terms of gantry control accuracy was also assessed via an analysis of the angular separation between adjacent projections. The median CNR increased linearly from 5.90 (20 projections/bin) to 8.39 (40 projections/bin). The ERW did not significantly change from 1.08 mm (20 projections/bin) to 1.07 mm (40 projections/bin), indicating the sharpness is not dependent on the total number of projections acquired. Scan times increased with increasing total projections and slower breathing rates. Across all 40 RMG-4DCBCT scans performed, the average difference in the acquired and desired angular separation between projections was 0.64°. RMG-4DCBCT provides the opportunity to enable fast low-dose 4DCBCT (∼70 s, 200 projections), without compromising on current clinical image quality.</jats:p>
Palabras clave: General Nursing.
Pp. 025009
EPOC outside the shield: comparing the performance of a consumer-grade EEG device in shielded and unshielded environments
Jordan Wehrman; Sidsel Sörensen; Peter de Lissa; Nicholas A Badcock
<jats:title>Abstract</jats:title> <jats:p>Low-cost, portable electroencephalography (EEG) devices have become commercially available in the last 10 years. One such system, Emotiv’s EPOC, has been modified to allow event-related potential (ERP) research. Although the EPOC has been shown to provide data comparable to research-grade equipment and has been used in real-world settings, how EPOC performs without the electrical shielding, commonly used in research-grade laboratories, is yet to be systematically tested. In the current article we address this gap by conducting a simple EEG experiment in shielded and unshielded contexts. Participants (n = 13, mean age = 23.2 years, SD = 7.9) monitored the presentation of human versus wristwatch faces, responding whether the images were inverted or not. This method elicited the face-sensitive N170 ERP. In both shielded and unshielded contexts, the N170 amplitude was larger when participants viewed human faces and peaked later when a human face was inverted. More importantly, Bayesian analysis showed no difference in the N170 measured in the shielded and unshielded contexts. Further, the signal recorded in both contexts was highly correlated. The EPOC appears to reliably record EEG signals without a purpose-built electrically-shielded room.</jats:p>
Palabras clave: General Nursing.
Pp. 025010
Analysis of the validity of the mathematical assumptions of electrical impedance tomography for human head tissues
Toby Williams; Kaddour Bouazza-Marouf; Massimiliano Zecca; Alexander L Green
<jats:title>Abstract</jats:title> <jats:p> <jats:italic>Objective:</jats:italic> To determine the validity of the key mathematical assumptions used in electrical impedance tomography for human head tissues. <jats:italic>Approach:</jats:italic>Conductivity and permittivity data collected from available literature for each tissue within the human head have been evaluated and critiqued. The most relevant dielectric tissue data for each tissue was then used to assess the validity of the mathematical assumptions of electrical impedance tomography in terms of their suitability for human head imaging in order to estimate related errors. <jats:italic>Main Results:</jats:italic>For induced currents with frequencies greater than 200 Hz the internal current source density is negligible. The assumption that magnetic effects are negligible is valid to an error of 1.7% for human head tissues for frequencies below 1 MHz. The capacitive effects are negligible for CSF, dura mater, blood, bone (cortical), and deep tissue skin for frequencies less than 3.2 MHz, 320 kHz, 25 kHz, 3.2 kHz, and 130 Hz respectively. However, the capacitive effects are not negligible for brain tissues, as the minimum error for brain tissues across the frequency range of 10 Hz to 100 GHz is 6.2% at 800 Hz, and the maximum error is 410% at 20 GHz. <jats:italic>Significance:</jats:italic>It is often assumed that the mathematical reduction of the base equations is valid for human head tissues over a broad frequency range; this study shows that these assumptions are not true for all tissues at all frequencies. False assumptions will result in greater errors and local distortions within tomographic images of the human head using electrical impedance tomography. This study provides the relationships between injected current frequency and the validity of the mathematical assumptions for each individual tissue, providing greater awareness of the magnitude of possible distortions.</jats:p>
Palabras clave: General Nursing.
Pp. 025011
In-line MRI-LINAC depth dose measurements using an in-house plastic scintillation dosimeter
Levi Madden; Natalia Roberts; Urszula Jelen; Bin Dong; Lois Holloway; Peter Metcalfe; Anatoly Rosenfeld; Enbang Li
<jats:title>Abstract</jats:title> <jats:p>Plastic scintillation dosimeters (PSDs) have many properties that make them desirable for relative dosimetry with MRI-LINACs. An in-house PSD, Farmer ionisation chamber and Gafchromic EBT3 film were used to measure central axis percentage depth dose distributions (PDDs) at the Australian MRI-LINAC Mean errors were calculated between each detector’s responses, where the in-house PSD was on average within 0.7% of the Farmer chamber and 1.4% of film, while the Farmer chamber and film were on average within 1.1% of each other. However, the PSD systematically over-estimated the dose as depth increased, approaching a maximum overestimation of the order of 3.5% for the smallest field size measured. This trend was statistically insignificant for all other field sizes measured; further investigation is required to determine the source of this effect. The calculated values of mean absolute error are comparable to the those of trusted dosimeters reported in the literature. These mean absolute errors, and the ubiquity of desirable dosimetric qualities inherent to PSDs suggest that PSDs in general are accurate for relative dosimetry with the MRI-LINAC. Further investigation is required into the source of the reported systematic trends dependent on field-size and depth of measurement.</jats:p>
Palabras clave: General Nursing.
Pp. 025012
Measuring prompt gamma-ray emissions from elements found in tissue during passive-beam proton therapy
Jeyasingam Jeyasugiththan; Jaime Nieto Camero; Julyan Symons; Pete Jones; Andy Buffler; Dieter Geduld; Stephen W Peterson
<jats:title>Abstract</jats:title> <jats:p>Prompt gamma detection during proton radiotherapy for range verification purposes will need to operate in both active and passive treatment beam environments. This paper describes prompt gamma measurements using a high resolution 2″ × 2″ LaBr<jats:sub>3</jats:sub> detector for a 200 MeV clinical passive-scatter proton beam. These measurements examine the most likely discrete prompt gamma rays emitted from tissue by detecting gammas produced in water, Perspex, carbon and liquid-nitrogen targets. Measurements were carried out at several positions around the depth corresponding to the location of the Bragg peak for water and Perspex targets in order to investigate prompt gamma emission as a function of depth along the beam path. This work also focused on validating the Geant4 Monte Carlo model of the passive-scatter proton beam line and LaBr<jats:sub>3</jats:sub> detector by making a direct comparison between the simulated and experimental results. The initial prompt gamma measurements were overwhelmed by the high amount of scattered radiation when measuring at isocenter, shifting the target further downstream from the final collimator significantly reduced the background radiation. Prompt gamma peaks were then clearly identified for the water, Perspex and graphite targets. The developed Geant4 Monte Carlo model was able to replicate the measured prompt gamma ray energy spectra, including production for important photopeaks to within 10%, except for the 4.44 MeV peak from the water target, which had more than a 50% overestimation of the number of produced prompt gamma rays. The prompt gamma measurements at various depths correlated well with the proton dose deposition; the 4.44 and 6.13 MeV photopeak profiles peaked within 1 cm of the Bragg peak and the R<jats:sub>50%</jats:sub> value for the 3–7 MeV energy range predicted the proton range within 8 mm.</jats:p>
Palabras clave: General Nursing.
Pp. 025013
Denoising PET images for proton therapy using a residual U-net
Akira Sano; Teiji Nishio; Takamitsu Masuda; Kumiko Karasawa
<jats:title>Abstract</jats:title> <jats:p>The use of proton therapy has the advantage of high dose concentration as it is possible to concentrate the dose on the tumor while suppressing damage to the surrounding normal organs. However, the range uncertainty significantly affects the actual dose distribution in the vicinity of the proton range, limiting the benefit of proton therapy for reducing the dose to normal organs. By measuring the annihilation gamma rays from the produced positron emitters, it is possible to obtain a proton induced positron emission tomography (pPET) image according to the irradiation region of the proton beam. Smoothing with a Gaussian filter is generally used to denoise PET images; however, this approach lowers the spatial resolution. Furthermore, other conventional smoothing processing methods may deteriorate the steep region of the pPET images. In this study, we proposed a denoising method based on a Residual U-Net for pPET images. We conducted the Monte Carlo simulation and irradiation experiment on a human phantom to obtain pPET data. The accuracy of the range estimation and the image similarity were evaluated for pPET images using the Residual U-Net, a Gaussian filter, a median filter, the block-matching and 3D-filtering (BM3D), and a total variation (TV) filter. Usage of the Residual U-Net yielded effective results corresponding to the range estimation; however, the results of peak-signal-to-noise ratio were identical to those for the Gaussian filter, median filter, BM3D, and TV filter. The proposed method can contribute to improving the accuracy of treatment verification and shortening the PET measurement time.</jats:p>
Palabras clave: General Nursing.
Pp. 025014
Monte Carlo calculated detector-specific correction factors for Elekta radiosurgery cones
P S Renil Mon; V N Meena Devi; Saju Bhasi; Sneha S Nair
<jats:title>Abstract</jats:title> <jats:p>A radiation field is considered small if its dimension is lower than the range of secondary electrons and the collimating devices partially occlude the source. Different detector types, such as unshielded diodes, diamond detectors, and small-volume ion chambers, are used for small-field measurements. Although the active volumes of these detectors are small, their non-water equivalent materials cause response variations. Herein, we aim to calculate the correction factors for our clinical detectors, EDGE detector (Sun Nuclear), 60017 diode (PTW), and CC01 ion chamber (IBA), for stereotactic radiosurgery cones of diameters of 5–15 mm in an Elekta Synergy linear accelerator using a Monte Carlo simulation. An Elekta Synergy linear accelerator treatment head was simulated using BEAMnrc Monte Carlo code as per the manufacturer specification. All three detectors were simulated as per the manufacturer specification. Three EGSnrc user codes were used for the detector simulation based on the detector geometry. The Monte Carlo model of the treatment head was validated against the measured data for a standard field size of 10 × 10 cm<jats:sup>2</jats:sup>. The off-axis profile, percentage depth dose, and tissue phantom ratio <jats:inline-formula> <jats:tex-math> <?CDATA ${{\rm{TPR}}}_{10}^{20}$?> </jats:tex-math> <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" overflow="scroll"> <mml:msubsup> <mml:mrow> <mml:mi mathvariant="normal">TPR</mml:mi> </mml:mrow> <mml:mrow> <mml:mn>10</mml:mn> </mml:mrow> <mml:mrow> <mml:mn>20</mml:mn> </mml:mrow> </mml:msubsup> </mml:math> <jats:inline-graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="bpexabe2bbieqn1.gif" xlink:type="simple" /> </jats:inline-formula> were verified in the validation procedure. The measured and Monte Carlo calculated relative output factors (ROFs) were not consistent. In a 5 mm field size, EDGE diode overestimated the ROF by 7.06%, and 60017 diode to 4.611%. In a 7.5 mm field size, the variations were 4.295% and 3.691% for EDGE and 60017 diodes, respectively. CC01 ion chamber under-responded up to 10% because of its low-density active volume. The maximum corrections were obtained in the smallest field size, which were 0.939(0.007), 0.962(0.006), and 1.117(0.008) for EDGE, PTW T60017, and CC01 detectors, respectively. After applying the Monte Carlo calculated correction factor to the measured ROF, it became consistent with the Monte Carlo calculated ROF.</jats:p>
Palabras clave: General Nursing.
Pp. 025015
Arterial stiffness as a vascular contribution to cognitive impairment: a fluid dynamics perspective
Trevor Tucker
<jats:title>Abstract</jats:title> <jats:p>A model of cerebral pulsatile blood flow through multiple arterial bifurcations is developed, based on the physics of wave propagation in compliant vessels. The model identifies the conditions for the optimum antegrade flow of blood into the arterioles as a function of the areas and stiffnesses of the arteries. The model predicts and quantifies the reduction in vessel diameter which occurs in progressing from the large central arteries into the arterioles. It also predicts and quantifies the change in vessel compliance which occurs in progressing from the large central arteries, through the small arteries, into the arterioles. Physics predicts that the clinically observed compliance changes are consistent with the efficient delivery of blood to the cerebral capillary bed. The model predicts that increasing arterial stiffening with age, reduces pulsatile cerebral blood flow substantially, potentially resulting in ischemia, hypoperfusion and hypoxia, with attendant neurological and cognition consequences. The model predicts that while central pulse pressure increases with aging, small vessel pulse pressure reduces, contrary to the concept of a pressure wave tsunami in the small vessels. The model also predicts that increased luminal diameters with increasing age, mitigate, somewhat the negative consequences of arterial stiffening, a form of adaptive arterial remodelling.</jats:p>
Palabras clave: General Nursing.
Pp. 025016